Hanna Instruments Worldwide
The Hanna Advantage: A Worldwide Leader in Technology and Innovation 100% Quality  360 Value
Contact Hanna: +27 (0) 11 615-6076
Hanna Instruments Worldwide
Customer Support: +1 (800) 426-6287

ISE Slope Check Buffers with Chloride Background - HI4000-47

R2,570.00 Excl VAT

Availability on Request

HI4000-47 is a set of pH 4 and pH 7 buffer packets (includes 10 packets of each buffer). These buffer packets contain chloride ions to allow an inner electrode check of the ammonia and carbon dioxide ion selective electrodes.

The ammonia and carbon dioxide sensors are both gas sensing, combination ion selective electrodes (ISEs) that contain an internal glass pH element to determine ammonia and carbon dioxide concentrations, respectively. An inner electrode check with the HI4000-47 buffers ensures the pH element is in working condition, permitting accurate ISE measurements.

  • Prepared with high purity chemicals
  • Pre-measured buffers for ease of use
  • Clearly marked lot number and expiration date

Description

HI4000-47 is a package containing pH 4 and pH 7 buffer packets with chloride ions and two plastic beakers. The special pH 4 and pH 7 buffers have a chloride background to verify the inner pH glass assembly housed within the ammonia and carbon dioxide gas sensing ISEs.

Before assembling an ammonia or carbon dioxide ISE for the first time, or if reactivating after storage, the inner glass assembly should be conditioned and tested as a pH electrode. The pH buffer packets can be added to 50 mL of deionized or distilled water and the sensor assembly carefully immersed in solution. When the mV measurement in each buffer solution is stable and recorded, the mV difference can be calculated. A calculated mV difference equal or greater than 160 mV is acceptable for temperatures between 20 and 25℃.

Additional information

Weight 0.08 kg
Parameter

pH

Specifications

SKU HI4000-47
Name ISE Slope Check Buffers with Chloride Background – HI4000-47
Certificate of Analysis No
Package packets and beakers
Quantity 10 packets each buffer; 2 beakers

Manuals

ISE Slope Check Buffers with Chloride Background – HI4000-47