My Dissolved Oxygen electrode is dry. What should I do?

Remove the red and black plastic cap or the membrane assembly. Soak the bottom 1 inch in electrolyte solution for 5 minutes. Rinse the membrane with electrolyte and refill with clean electrolyte. Gently tap the sides of the membrane cap to ensure that no air bubbles remain trapped. Adjust O-ring inside membrane cap. With the sensor facing down screw the membrane assembly

Remove the red and black plastic cap or the membrane assembly. Soak the bottom 1 inch in electrolyte solution for 5 minutes. Rinse the membrane with electrolyte and refill with clean electrolyte.... read more »
Close window
My Dissolved Oxygen electrode is dry. What should I do?

Remove the red and black plastic cap or the membrane assembly. Soak the bottom 1 inch in electrolyte solution for 5 minutes. Rinse the membrane with electrolyte and refill with clean electrolyte. Gently tap the sides of the membrane cap to ensure that no air bubbles remain trapped. Adjust O-ring inside membrane cap. With the sensor facing down screw the membrane assembly

No results were found for the filter!
Viewed