Chlorine Portable Photometer (HI97790 / HI97790C)

R12,471.00R18,610.00 Excl VAT

The HI97790 meter measures free and total chlorine in water samples from 0.00 to 5.00 mg/L (ppm). The method is an adaptation of the Standard Methods, 4500‐Cl G, DPD Colorimetric Method.

Description

Description:
The HI97790 meter measures free and total chlorine in water samples from 0.00 to 5.00 mg/L (ppm). The method is an adaptation of the Standard Methods, 4500‐Cl G, DPD Colorimetric Method. Chlorine is a widely used disinfectant in the treatment of drinking water, wastewater and swimming pools.

Main Features:
• Only one sample needed for successive free and total chlorine measurements.
– Take a ZERO reading
– Add required reagents for Free Chlorine testing
– Read Free Chlorine levels
– Add required reagents for Total Chlorine testing
– Read Total Chlorine levels
• Ideal for regulatory compliance in areas requiring successive sample measurements.
• Easy-to-use and cost effective while maintaining high accuracy
• Advanced LED optical system
– Innovative optical design that utilizes a reference detector and focusing lens to eliminate errors from changes in the light source and from imperfections in the glass cuvette.
– LEDs have a much higher luminous efficiency, providing more light while using less power. They also produce little heat, which could otherwise affect electronic stability.
– The optical system is sealed from outside dust, dirt and water.
– An exclusive positive‐locking system ensures the cuvettes are placed into the holder in the same position every time.
• CAL Check™
– Validate instrument performance at any time using CAL Check cuvettes made with NIST traceable standards. The CAL Check screen guides the user step-by-step through the validation process and user calibration.
• On-screen tutorial mode with animations
– Guides users step-by-step through the measurement process
• Waterproof and floating IP67 case
• Unit of measure is displayed along with reading
• Built-in timer- Built-in reaction timer that ensures consistency between tests.
• Error messages on display
– Alerts to problems including no cap, high zero, and standard too low
• GLP data:
– Displays the last calibration date.
• Auto data log and recall up to 50 individual measurements
• Battery status indicator
• Auto-shut off

Primary Applications:
Drinking water, wastewater and swimming pools and spas

 

Additional information

Weight 0.380 kg
Dimensions 14.25 × 10.25 × 5.05 cm
Available in a Kit or Meter Only

"Meter Only", "Kit"

Specificiations

Specifications:

Chlorine (all methods) Range 0.00 to 5.00 mg/L (as Cl₂)
Resolution 0.01 mg/L
Accuracy ±0.03 mg/L ±3% of reading at 25 °C
Method Adaptation of Standard Methods, 4500-Cl G, DPD Colorimetric Method
Measurement system Light source Light emitting diode
Bandpass filter 525 nm
Bandpass filter bandwidth 8 nm
Bandpass filter wavelength accuracy ±1.0 nm
Light Detector Silicon photocell
Cuvette type Round 24.6 mm diameter (22 mm inside)
Additional specifications Auto logging 50 readings
Display 128 x 64 pixel B/W LCD with backlight
Auto-off After 15 minutes of inactivity ((30 minutes before a Free Cl2 / Total Cl2 measurement)
Battery type Alkaline 1.5 V AA (3)
Battery Life > 800 measurements (without backlight)
Environment 0 to 50 °C (32 to 122 °F); 0 to 100% RH, non-

serviceable

Dimensions 142.5 x 102.5 x 50.5 mm (5.6 x 4.0 x 2.0″)
Weight 380 g (13.4 oz.); with batteries
Casing IP67, floating case

Ordering Information

HI97790 is supplied with sample cuvette (2 pcs.); sample cuvette cap (2 pcs.); plastic stopper (2 pcs.); 1.5V AA alkaline battery (3 pcs.); quick reference guide with instructions for manual download and instrument quality certificate.
CAL Check standards and testing reagents sold separately

HI97790C includes photometer; rugged carrying case; sample cuvette (2 pcs.); sample cuvette cap (2 pcs.); plastic stopper (2 pcs.); A ZERO – CAL Check™ cuvette A; HI97701B – CAL Check cuvette B for Free and Total Chlorine; cloth for wiping cuvettes; scissors; 1.5V AA alkaline battery (3 pcs.); CAL Check standard certificate; quick reference guide with instructions for manual download and instrument quality certificate.

Reagents sold separately

Accessories:
HI97701-11 CAL Check™ standards for Free and Total Chlorine -cuvette kit
HI93701-F free chlorine liquid reagent for 300 tests
HI93701-T total chlorine liquid reagent for 300 tests
HI7101412 HI97 series blue carrying case with 2 cuvette slots
HI731318 Cloth for wiping cuvettes (4 pcs.)
HI731331 Glass cuvette (4 pcs.)
HI731336N Cap for glass cuvette (4 pcs.)
HI93703-50 Cuvette cleaning solution (250 mL)

Download Manual