Coulometric Karl Fischer Titrator, With Diaphragm (HI934D-02)

R253,524.00 Excl VAT

SKU: HI934D-02 Categories ,

Out of stock

For scientists and professionals who need exact water content determination from 1 ppm to 5%, our new generation of the Coulometric Karl Fischer Titrator is engineered to meet or exceed your technical requirements while delivering a lower cost of ownership. With cutting-edge features packed into a compact design, this titrator delivers accurate results and requires less space in the lab. Get accurate water content determination at the push of a button. Designed for the modern lab, this karl fischer titrator is the perfect fit for your testing environment.

      • Compact design will save space in the lab without compromising your results.

 

      • Durable heat and chemical resistant body.

 

    • Sealed solvent system helps to keep water out of the system, while giving you minimal exposure to the reagents.

Description

Titrator Capabilities

Dynamic Titrant Dosing

The dynamic titration speed allows for timely and accurate titration results by relating the amount of iodine generated to the mV response from the Karl Fischer reaction. This provides for a larger generation of iodine near the beginning of a titration and more precise pulses near the titration endpoint.

Drift Rate Compensation

The HI934 automatically adjusts the titration calculation to account for the effects of any ambient humidity entering the titration cell. This provides a more accurate result by correcting for water not present in the actual sample.

Selectable Endpoint Criteria

The HI934 employs a dual platinum pin electrode for bivoltammetric endpoint determination. Users may choose termination criteria based on mV stability times or drift rates.

Multistage Cell Preparation

A pre-titration stage eliminates residual water present in the solvent and the cell, providing a reliable baseline start to analysis. Standby mode then keeps the solvent dry between titrations and when the titrator is not in use.

Coulometric Titration System

Purposeful Generator Design

Two styles of generators are available for the HI934; a fritted design ideal for samples with low water content and high accuracy demand; and a fritless generator allowing for easier cleaning and maintenance.

Precision Iodine Generation

Our dosing algorithm allows for an extremely small amount of iodine necessary for the Karl Fischer reaction to be generated electrolytically using a pulsed current up to 400 mA, delivering iodine accurately and precisely.

Titration & Solvent System

Chemically Resistant Vessel & Tubing

The glass titration cell and PTFE tubing is designed to withstand the harsh solvents and reagents involved in Karl Fischer reactions.

Sealed Solvent System

Ground glass joints completely seal the glass titration cell, minimizing exposure to ambient humidity, keeping the system dry, and reducing reagent consumption while saving time between titrations. Solvent may be exchanged in a matter of seconds with a quick fitting adjustment.

High Efficiency Desiccant

High efficiency molecular sieve desiccant helps maintain low and stable drift rates within the titration cell while preventing the ingress of ambient humidity into the sealed solvent system.

Interface & Display

Interactive Color Display

A large, color LCD screen clearly shows the chosen titration method along with results, units, total water, drift rate, and mV value.

Detailed Titration Graphs

A real-time titration curve can be displayed during each titration; this feature is useful when new methods are tested or when a procedure requires optimization.

Simple & Quick Navigation

Virtual key selections present on the display allow for simple and quick navigation between screens and menus without getting lost in a nest of information.

Data & Storage

USB Transfer

Customizable Titration Reports

Each titration report is fully customizable so users can ensure they are storing and filing the appropriate data required for their application and procedures.

Flexible GLP Management

All necessary GLP (Good Laboratory Practice) information can be recorded with each sample including: sample identification, company and operator name, date, time, and electrode ID codes.

Effortless Data Transfer

Data can easily be transferred to a USB flash drive or PC with the Hanna HI900PC application software. The USB port allows for the transfer of titration methods, titration reports, and software upgrades via USB flash drive.

Connectivity & Functionality

HI902C Angled

Configurable Balance Interface

Sample size may be automatically entered from any laboratory analytical balance with a RS232 serial output saving time and labor.

Multiple Peripherals

Users can print reports directly from the titrator using a standard parallel printer. An external monitor and keyboard may be attached for added versatility, as well as an analytical balance for automatic sample mass entry for titrations.

Methods of Analysis

Methods of Analysis

Customizable Methods

The HI934 can store up to 100 user-defined or standard titration methods. Each method may be customized and optimized for performance based on application and user requirements.

Titration Method Support

Onsite installation, training, and customization is available from one of our Applications or Service experts. Hanna offers continued support via phone or webinar for any questions you might have along the way.

Adaptable Standard Methods

Our technical experts can program and customize standard methods developed by such affiliations as ISO, ASTM, AOAC, AOCS, EPA, and more directly onto your titrator. Ask our Sales Consultants which standard methods are possible with our HI934 Karl Fischer system.

Additional information

Parameter

Titration

Specifications

Measurement
Range 1 ppm to 5%
Resolution 0.1 ppm
Result Units % ppm, mg/g, µg/g, mg, µg, mg/mL, µg/mL, mgBr/100g, gBr/100g, mgBr, gBr
Sample type liquid or solid (external dissolution or extraction)
Determination
Pre Titration conditioning automatic
Background Drift Correction automatic or user selectable value
Endpoint Criteria fixed mV persistence, relative drift stop, or absolute drift stop
Dosing Dynamic with 3 speed settings
Result Statistic mean, standard deviation
Titration Vessel
Type borosilicate glass with standard taper glass joint connections
Operating volume 100 to 200 mL
Septum silicone rubber
Septum Cap Thread GL-18
Reagent Port Standard Taper 19
Detector Electrode
Type/Connection Dual platinum pin, polarisation electrode/BNC Connector
Glass connection standard Taper 14/20
Polarisation Current 1, 2, 5, or 10 µA
Voltage Range 2 to 1100mV
Voltage Resolution 0.1mV
Accuracy (@25°C) ±0.1%
Generator Electrode
Type diaphragm or diaphragm-less
Electrode Type Detection Automatic
Electrical Connection 5-pin connector with detachable cable
Glass Connection standard taper 29/12
Maximum Current 400 mA
Current Control automatic or fixed (400 mA)
Stirrer
Type Magnetic, electronic regulated, digital stirrer
Speed 200 to 2000 RPM
Resolution 100 RPM
External Stirrer 4-pin mini DIN connection allows for the control of an external stirring apparatus
Reagent Handling System
Type sealed system with integrated diaphragm air pump
Desiccant Type Molecular sieves
Bottle thread type GL-45
Glass connection standard taper 19 (using supplied adapter)
Reagent/Waste Tubing PTFE
Additional Specifications
Display 5.7” graphical colour display with backlight
Peripheral Devices PC (USB standard B), flash drive (USB standard A), analytical balance (DB-9 socket), printer (DB-25 socket), keyboard (6-pin mini DIN)
Languages English, Portuguese, Spanish and French
Power supply/power draw 100-240 VAC, 50/60 Hz/0.5 Amps
Enclosure material ABS/PC and stainless steel
Keypad polyester
Operating Environment 10 to 40°C; up to 80% RH
Storage environment -20 to 70°C; up to 95% RH
Dimensions 315 x 205 x 375 mm
Weight approx. 4.3kg with 1 pump, stirrer and sensors
Ordering information HI-934D-01 and HI-934D-02 are supplied with diaphragm
HI-934-01 and HI-934-02 are supplied without diaphragm
All models include: Dual platinum pin electrode, air pump/stirrer assembly, titration vessel assembly (glass vessel, accessory port stopper, sample port cap and septum, stir bar, desiccant, desiccant cartridge fittings), vessel support with adapter, pump locking screw with plastic head, reagent bottle assembly (bottle cap, desiccant, desiccant cartridge, fittings, tubing – silicone and PTFE), water bottle assembly (waste bottle, bottle cap, desiccant, desiccant cartridge, fittings, tubing – silicone and PTFE), calibration key, reagent exchange adapter, accessory holder assembly, joint grease, Karl Fischer generator electrode (removable generator electrode cable), USB cable, USB storage device, HI-900 PC application software, power adapter, quality certificate and instruction manual binder.