Description
The BL100 uses the HI10053 amplified probe that incorporates both pH and temperature sensors and connects to the controller with a single waterproof Quick Connect DIN connector.
Built-in Amplifier
- The built-in amplifier helps to reduce electrical noise from recirculation pumps to provide for stable, reliable measurement.
pH Glass
- The pH glass used is ideal for low conductivity water and provides for a fast response.
PVDF Body
- The PVDF body of the probe has a 1/2” threaded fitting for insertion to an in-line “T” fitting or the flow cell. The probe body has a hex fitting for tightening snuggly with a wrench.
Safety Features
- The HI10053 is designed to have a 0 mV value at pH 4, and not at pH 7 like typical pH electrodes. In the event of a break, this reading will produce an alarm state (when programmed) and disable the pump from dosing. A standard pH electrode would have a pH value close to 7 and could result in either no dosing or excessive dosing based on the actual value and programmed set point.
Specifications
Specifications BL100
pH
Range*; 0.00 to 14.00 pH
Resolution; 0.01 pH
Accuracy (@25°C/77°F) ;±0.10 pH
Calibration; user calibration: automatic, one or two-point with buffer solution (4.01, 7.01, 10.01 pH)
Process calibration: single point, adjustable (± 0.50 pH around measured pH)
Temperature
Range* -5.0 to 105°C (23.0 to 221.0°F)
Resolution 0.1°C (0.1°F)
Accuracy (@25°C/77°F) ±0.5°C (±0.9°F)
Additional
Specifications
Temperature Compensation automatic
Dosing Control Type On/Off control using adjustable set point (6.00 to 8.00 pH) with adjustable hysteresis (0.10 to 1.00 pH)
Proportional control using adjustable set point (6.00 – 8.00 pH) with adjustable proportional band (0.10 to 2.00 pH)
Dosing Control Activation
high or low mode operation
high set point dosing is activated when reading is higher than set point (dose acid)
low set point dosing is activated when reading is lower than set point (dose base)
Delay Start for Dosing startup delay timer at power-on (0 to 600 sec.)
Maximum Dosing Time overfeed protection using overtime safety timer (1 to 180 min. or Off)
Pump Flow Control selectable flow rate (0.5 to 3.5 L / hour; 0.13 to 0.92 G/hour)
manual control for pump priming
Alarms
high and low with enable / disable option
triggered after 5 sec. if controller records a set of consecutive readings over / under threshold values
level with enable / disable option
overtime protection (1 to 180 min. or off)
intuitive alarm system using red, light green, and green color-coded backlight
Alarm Relay Output (1) SPDT 2.5A / 230 VAC
External Event Input input for level controller or flow switch to disable dosing pump in the event of no chemical when using a level controller
or no flow when using a flow switch – galvanically isolated
Probe Input (1) HI10053 digital pH/temperature probe with quick connect DIN connector – galvanically isolated
Power Supply 100—240 VAC, 50/60 Hz
Power Consumption 15 VA
Environment 0-50°C (32-122°F), max. 95% RH non-condensing
Dimensions 90 x 142 x 80 mm (3.5 x 5.6 x 1.8”)
Weight 910 g (32 oz.)
Casing wall mounted, built-in pump, IP65 rated