Description
Hanna’s EPA compliant turbidity meters are designed to meet or exceed the criteria specified by the USEPA Method 180.1 and Standard Method 2130 B. The USEPA Method 180.1 specifies the key parameters for the optical system to measure turbidity for drinking, saline, and surface water by use of the nephelometric method.
In the nephelometric method, the light beam that passes through the sample is scattered in all directions. The intensity and pattern of the scattered light is affected by many variables, such as wavelength of the incident light, particle size and shape, refractive index, and color. The optical system includes a tungsten replacement lamp, a scattered light detector (90°) and a transmitted light detector (180°).
The HI740234 is used with the EPA compliant turbidity meters including:
HI83414 Turbidity and Free/Total Chlorine Benchtop Meter
HI88703 Precision Benchtop Turbidity Meter
HI93414 Turbidity and Free/Total Chlorine Portable Meter
HI98703 Portable Turbidity Meter